Chemical Monitoring Testing Package Application for High Pressure-steam Sterilization:An Observation and Analysis
- VernacularTitle:压力蒸汽灭菌化学监测测试包的使用观察与分析
- Author:
Hua JIANG
;
Yunna QU
;
Fen AI
;
Guodong REN
;
Jianqiu YANG
- Publication Type:Journal Article
- Keywords:
Pressure-steam sterilization;
Chemical indicator card in package;
Monitoring efficacy
- From:
Chinese Journal of Nosocomiology
2006;0(12):-
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVE To observe and compare the application result of two kinds of chemical indicator(CI)card contained in high pressure-steam sterilization chemical monitoring testing package for providing correct credible evidence of supplying material in each batch after high pressure-steam sterilization and to avoid resource lost because of fault estimation.METHODS The chemical monitoring testing package was made according Sterilization Criteria published in 2002,with the 3M 1250 and 1243 CI cards and 1292 biology indicator(BI) to observe the results in tested package after sterilization.RESULTS The BI in chemical monitoring testing package was all qualified,the qualified rate of 1250 was 70%,while of 1243 was 100%.CONCLUSIONS If there is no BI for monitoring,1243 CI card should be chosen in high pressure-steam sterilization chemical monitoring testing package for supplying material in each batch after high pressure-steam sterilization which is not influenced by moisture and the contacted material and is easy to read and evaluate,the resource lost caused by fault estimation can be avoided.