BacT/Alert 3D Automatic Blood Culture System:Clinical Application and Evaluation
- VernacularTitle:BacT/Alert 3D血培养仪临床应用评价
- Author:
Chengbin ZHU
;
Lu DOU
;
Yongxiang XIA
- Publication Type:Journal Article
- Keywords:
Automatic blood culture system;
Bacteria culture;
Septicemia
- From:
Chinese Journal of Nosocomiology
2006;0(08):-
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVE To evaluate the clinical application of BacT/Alert 3D automatic blood culture system.METHODS Totally 3610 blood specimens were cultured by BacT/Alert 3D automatic blood culture system.RESULTS Four hundred and three of 3610 blood specimens were positive,and the positive rate was 11.2%.The cumulative positive rates within 24 h,24-48 h,48-72 h,and longer than 72 h were 64.3%,21.1%,11.4% and 3.2%,respectively.CONCLUSIONS BacT/Alert 3D automatic blood culture system is an accurate,quick and convenient method for bacteria culture,which can detect wide range of bacteria strains,elevate detection ratc,and shorten detection period.