Error Analysis of X—knife Treatment Plan
- VernacularTitle:X-刀治疗计划验证及误差分析
- Author:
Pinting ZHAO
;
Jun LIANG
;
Qiuju SHAO
;
Fang LIANG
;
Huiqin YUANG
;
Yongxin YIN
- Publication Type:Journal Article
- Keywords:
stereotactic;
radiotherapy;
errors
- From:
Chinese Medical Equipment Journal
1989;0(03):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To investigate the probability of verifying the X-knife treatment plan by the radiotherapy simulator,and to analyze and report the errors of the X-knife treatment plan for reference.Methods The radiation fields of the X-knife treatment plan were observed in the whole simulating treatment process by the radiotherapy simulator.Results The error rate of X-knife treatment plan were in the simulating treatment process was 5.9%(22/372).Among the errors,3 cases of X-knife plans were found in the simulating treatment process,and the errors of isocenter in Z direction were 10cm.Conclusion This is a special error of X-knife TPS,which deserves more attention in stereotactic radiotherapy.