Study on measuring method for amplitude of transducer’s tip in ultrasonic surgical apparatus
- VernacularTitle:超声外科手术装置检定系统中超声刀头振幅测量方法的研究
- Author:
Tiebing LIU
;
Liming TANG
;
Min WU
;
Shitao QI
- Publication Type:Journal Article
- Keywords:
ultrasonic surgical apparatus;
ultrasonic harmonic scalpel(UHS);
amplitude;
image capture;
digital image processing
- From:
Chinese Medical Equipment Journal
1989;0(01):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To study the measuring method for amplitude of transducer's tip in ultrasonic surgical apparatus, and design a set of practical detecting system with special data processing technique. Methods The system was consisted of stereomicroscope, CCD camera, etc. The images of amplitude measuring were captured and processed and the result was carried out at last. Results The measurement range for amplitude of transducer’s tip was above 190 ?m, the precision was below 10?m, and the repeatability was also ideal. Conclusion The amplitude measuring method has some excellent characters such as directness, simplicity, easy implementation, error estimableness and controllability, etc. It is proved in practice that the effect is nice and the blank in this field in China is filled.