Design of new type thermal container for infant
- VernacularTitle:新型婴儿保温箱的设计
- Author:
Huiqin SUN
;
Zhilan LIU
;
Nanping LIU
;
Xin LI
;
Feng LI
- Publication Type:Journal Article
- Keywords:
double temperature controller;
protection circuit;
database;
Single Chip Microcomputer Cygnal F005
- From:
Chinese Medical Equipment Journal
1989;0(01):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To develop infant thermal container with forced protection circuit so as to ensure its security. Methods Temperature database was set in thermal container and Cygnal F005 was used as intelligent controller. The state of the sensor was judged by comparison precisely. Double temperature controller was adopted to ensure normal working temperature. When the container went wrong, it could automatically switch to protection circuit to keep the temperature for infant. Results With precise temperature control and steady working station, the new type thermal container for infant could provide a safe and comfortable condition with clean air and fitting temperature and humidity. Conclusion The container is steady, accurate, safe and convenient.