Secure design of LEEP electro-scalpel
- VernacularTitle:LEEP刀的安全性设计
- Author:
Jingyuan ZENG
- Publication Type:Journal Article
- Keywords:
LEEP;
security;
leakage current;
power control
- From:
Chinese Medical Equipment Journal
2003;0(12):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To discuss how to deal with the problems of LEEP electro-scalpel that are caused by the burn of negative electrode loose and the electric shock of leakage current.Methods The system's security was enhanced by insulation,power and software control and it was also evaluated.Results The electric leakage of LEEP electro-scalpel was acceptable and the preventive support of burn was effective.Conclusion It is proved by the synthetical evaluation that the device's secure design is reliable.