Study and design of portable medical drop speed monitoring equipment
- VernacularTitle:医用便携式点滴监护装置研究
- Author:
Weidong SONG
;
Wanjun LEI
;
Haichao ZHANG
;
Bo SHI
- Publication Type:Journal Article
- Keywords:
photo-electric detect;
drop speed;
MCU
- From:
Chinese Medical Equipment Journal
2003;0(11):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To develop a portable medical drop speed monitoring equipment which can conveniently control the drop speed in infusion.Methods Taking the monolithic integrated circuit of 89C51 as a core,such units were designed as the master control unit,keyboard input system,drop speed detection,drop speed control,digital display,acousto-optics alarm circuit and electromotor drive.All circuit module units were connected in monolithic integrated circuit.The gathering data were transferred to monolithic integrated circuit in the form of electrical signals.After the operation,analysis and processing,monolithic integrated circuit transferred the data to demonstration module,electrical machinery and acousto-optics alarm circuit through the outlet.Results Automatically controlling the drop speed according to the difference of drop speed,this equipment can reach the required drop speed in 5~10s and maintain this speed with an error of 12 drops per minutes.Conclusion Portable medical drop speed monitoring equipment achieves the goal of the task and meets the clinical requirement.This equipment is worthy of popularizing.