Certification of linear range of WBC measured by CD3700 and application in ISO15189
- VernacularTitle:雅培CD3700测定白细胞线性范围的验证与ISO15189应用
- Author:
Youye YANG
;
Zhizhao YANG
;
Lishao MIAO
;
Fuda HUANG
;
Jinsheng LIANG
- Publication Type:Journal Article
- Keywords:
white blood cell;
linear range;
certification
- From:
Basic & Clinical Medicine
2006;0(10):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To certify the linear range of WBC measured by CD3700.Methods Use the white stratum of blood after centrifugation.Then,the high value specimens were diluted with saline to get a series of samples with different concentrations.Every specimen was measured twice in order to keep the sequence of specimens randomly.The perform outlier test,polynomial regression analysis,and the random error estimation were carried out.Results There was no outlier in the results.It was demonstrated by polynomial regression to be linear from(0~219)?109 L-1.The random error was within the allowable limit of 1/4CLIA'88(3.75%).Conclusion The linear range of WBC measured by CD3700 is just like that of the manufacturer claims.