Electrical Impedance Tomography System for lung function monitoring
- VernacularTitle:用于肺功能监测的电阻抗成像系统
- Author:
Huaxiang WANG
;
Lu LI
;
Chunyan WANG
- Publication Type:Journal Article
- Keywords:
Electrical Impedance Tomography;
DDS;
phase-sensibility demodulation
- From:
Chinese Medical Equipment Journal
2004;0(07):-
- CountryChina
- Language:Chinese
-
Abstract:
A complete Electrical Impedance Tomography system that can monitor lung function is introduced in this paper. The master PC is responsible for visual control and image reconstruction, while DSP carries out tasks such as electrode selection, data acquisition and data communication with the master. Based on the DDS technique, the exciting signal source has high accuracy, and the data can be acquired by phase-sensibility demodulation. The vivo experiment proves that gravity exerts an effect on the distribution of intrapulmonary ventilation, and shows that the Electrical Impedance Tomography is able to monitor lung circulation and check pathological changes.