Circuit analysis and PCB adjusting of R-500 X-ray machine
- VernacularTitle:R-500型X线机电路解析与PCB板调节
- Author:
Fanglin ZHANG
;
Qinghua WANG
;
Xiangyang WU
;
Yadong LI
- Publication Type:Journal Article
- Keywords:
electrical circuit analysis;
indicator light;
PCB adjusting
- From:
Chinese Medical Equipment Journal
1989;0(02):-
- CountryChina
- Language:Chinese
-
Abstract:
Abstract PCB faults on X-ray machine occur frequently in our hospital due to humid climate, which lead to many malfunctions such as potentiometer failure, blank screen, lacking KV display, unavailable parameter choosing and disappearing of Er1~Er7 fault code. These malfunctions could be diagnosed and repaired quickly based on analyzing the status of the machine's workflow indicator lights, as well as studying circuit and function principle of the equipment.