Development and clinical application of Heat Probe(HP-A) in the endoscopic treatment
- VernacularTitle:HP-A内镜热极治疗仪的研制和临床应用
- Author:
Xiaoping WANG
;
Jian JIANG
;
Maohai SONG
;
Chengrong SONG
;
Shuhu WU
- Publication Type:Journal Article
- Keywords:
Heat Probe;
endoscopic therapy;
Thermocoagulation;
compatibility
- From:
Chinese Medical Equipment Journal
2004;0(07):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To develop a portable endoscopic thermal device which possesses the merits of compact structure,convenient operation,security and good compatibility.Methods Rapid heating elements and several materials are used,which are excellent insulation,heat conduction and can withstand high temperatures.The production process of miniaturization,anti-conglutination and constant temperatureare adopted.Results: The size of Heat probe is about 2.4mm in outer diameter and 8 mm in length.The maximum temperature will reach 250℃,heating time within 10s,output 4.5~6.5W.Conclusion Heat Probe can be compatible with various types gastrointestinal endoscopic channel.Besides hemostasis of upper gastrointestinal,it can also be applied to small polyps,erosive gastritis,etc.