Monitoring of sterilization quality of pulsating vacuum sterilizer in our hospital
- VernacularTitle:我院脉动真空灭菌器灭菌质量的监测
- Author:
Lizhong DAI
;
Baixiang ZHENG
- Publication Type:Journal Article
- Keywords:
pulsating vacuum sterilizer;
sterilizing quality;
monitoring
- From:
Chinese Medical Equipment Journal
2004;0(07):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To find out sterilization quality of pulsating vacuum sterilizer for better sterilization effect.Methods The sterilization quality was monitored by physical,chemical and biological methods.Results 4,576 subjects were involved in the test.Such data were obtained as the qualification rate for temperature,the one by chemical indicator card,the one by B-D test and the one by biological indicator,which were 99.6%,99.8%,99.9% and 100% respectively.Conclusion Above data proves that the sterilizer is in normal condition.