Study on repacking of HITACHI SX-YB2 X-ray photography system
- VernacularTitle:日立SX—YB_2型X线摄影系统改装的研究
- Author:
Yanjun LU
;
Lishan KUANG
;
Ming LI
;
Ling MOU
;
Bingxiao FENG
- Publication Type:Journal Article
- Keywords:
HITACHI;
indigenous X-ray;
repark
- From:
Chinese Medical Equipment Journal
1993;0(06):-
- CountryChina
- Language:Chinese
-
Abstract:
The new general photography system is composed of overhang of SX-YB2 departed from the mainframe of HITACHI Digital Fluoroscpic(model 110XF),topmanagement system of FSK302-1 and FSB302-1 and leveled adiogaphy system of F78-III.The filament heater circuit,rotating anode startup and running circui and X-ray produce circuit are mainly changed,and some circuits in power and filament heat are added.