Automated capillary zone electrophoresis for analysis of human serum protein in clinical application
- VernacularTitle:全自动多通道毛细管区带电泳法在血清蛋白分析中的临床应用
- Author:
Yan CHEN
;
Zeyu HUANG
;
Yajie WANG
;
Suping WANG
;
Xixiong KANG
- Publication Type:Journal Article
- Keywords:
Blood protein electrophoresis;
Electrophoresis, agargel;
Electrophoresis,capillary
- From:
Chinese Journal of Laboratory Medicine
2001;0(05):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective The automated capillary zone electrophoresis (CE) is used for human serum protein analysis and compared with the results of agarose gel electrophoresis (AGE) for human serum protein analysis.Discuss its′ practical application in clinical.Methods With the Capillarys ?1-?2+reagent set, proteins in serum were separated at 7 kV for 4 min in 15.5 cm?25.0 ?m fused-silica capillaries (n=8) at 35.5℃ in a pH 10 buffer with online detection 200 nm.Results Capillary zone electrophoresis compared with agarose gel electrophoresis(AGE;Hydrasys-Hyry; Hydragel protein(e)15/30/54 reagent set;Sebia)in precision, relatives, linearity and potential interferences were analyzed.A samples without Para-protein (n=165) were performed relation test. The coefficient of relation between CE and AGE is 0.929 for albumin, 0.924 for for ?1-globulin, 0.841 for ?2-globulin, 0.789 for ?-globulin, 0.926 for ?-globulin. There was no significant difference (P