A comparative study of platelet rich plasma stored in two kinds of platelet incubators
- VernacularTitle:两种血小板保存箱保存富血小板血浆效果的比较
- Author:
Wei HAN
;
Jinghan LIU
;
Xilin OUYANG
- Publication Type:Journal Article
- Keywords:
Platelet preservation;
Comparison;
CD62p
- From:
Chinese Journal of Blood Transfusion
2002;0(05):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To study the preservative indices of stored platelet and verify the effect of a new platelet incubator made in China. Methods 12 bags of fresh platelet rich plasma (PRP) were collected. Each bag was equally divided into 2.One bag was stored for 5 days in XHZ-IA platelet incubator which was made in China, and the other bag stored in FORMA-3606 platelet incubator made in USA. Some preservative indices of stored platelet e.g. aggregation rate, recovery rate after response to hypotonic shock, CD62p expression rate etc. were tested daily. Results There were no significant differences in these preservative indices between platelets stored in two platelet incubatorshas. Conclusion XHZ-IA platelet incubator has the same effect on PRP preservation as FORMA-3606 platelet incubator.