Twolayer structure of the hollow stent wall for bloodvessel anastomosis
- VernacularTitle:血管吻合用中空支架壁双层结构的研究
- Author:
Geng SUN
;
Boxun ZHANG
;
Xiangdang LIANG
;
Yan WANG
- Publication Type:Journal Article
- Keywords:
twolayer stent;
gelatin;
polyglycol 20000
- From:
Chinese Medical Equipment Journal
1993;0(05):-
- CountryChina
- Language:Chinese
-
Abstract:
A stent for bloodvessel anastomosis, with a slippery, soft out layer and a rapidly soluble inner layer, is developed in this paper. The out layer is made up of the gelatin with a high purity and a low molecular weight, and the inner one of polyglycol with a molecular weight of 20000. The structure changes of the stent surface are studied with anastomosis environment simulated. The result proves that the stent can be applied to bloodvessel anastomosis