Comparison and evaluation study on two hepatitis B virus DNA quantitative tests
- VernacularTitle:两种乙型肝炎病毒DNA定量检测方法的比较与评价
- Author:
Hao WANG
;
Qimin TAO
;
Juan WU
;
Li ZHANG
;
Yan SUN
- Publication Type:Journal Article
- Keywords:
Hepatitis B virus;
DNA;
quantitative test
- From:
Chinese Journal of Laboratory Medicine
2001;0(05):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To evaluate the two methods for quantitative hepatitis B virus (HBV) DNA test. Methods The Hybrid Capture II system from Digene Co. and Real Time PCR fluorimetry quantitative HBV DNA test kit from PIJI Bio Technical Development Company Ltd. were used to detect sera HBV DNA in sera of patients with chronic hepatitis B. Results With the two quantitative methods, the positive rates of HBV DNA were 98 6% and 94.6% respectively. The concordant rate of these two methods was 93.2%. In assessing the sensitivity of the two kits with a serial of 10 fold diluted patients′s sera, PG kit could test 10 -7 of diluted serum, and HC II could test 10 -6 . The test value of HC II was more accurate in quantitation than PG kit especially in higher titer of HBV DNA. Whereas in the case of low titer of HBV DNA, the deviation of test value increased in both two methods. For monitoring the anti virus effect with quantitative HBV DNA in CHB patients, these two methods had same sensitivity, and the test value of HBV DNA had the same trend of change. Conclusion HC II system and PG test kit were sensitive and reliable, and they were worthy of monitoring the anti virus efficacy and of clinical practice.