Studies on standardization of serum enzyme determination
- VernacularTitle:血清酶测定标准化的实验研究
- Author:
Jian GUO
;
Qingtao WANG
;
Qing TONG
;
Zhenhua YANG
- Publication Type:Journal Article
- Keywords:
Enzyme;
Standardization;
Calibrator
- From:
Chinese Journal of Laboratory Medicine
2001;0(03):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective Investigating for the standardization of serum enzyme determination by using enzyme calibrator in lab tests Methods 150 laboratories were involved in this program. One enzyme calibrator and three patient samples were send to each participate lab The calibrator was tested using routine method and record the results The patient samples were analyzed before and after calibration Record the results All lab data was processed with computer Results Compare with the target values, the bias of alanine aminotransferase (ALT), aspartate aminotransferase (AST), amylase (AMY) and lactate dehydrogenase (LD) assays were 3 8%, -1 8%, 2 3% and -5 2% respectively, the bias of alkaline phosphatase (ALP), gamma glutamyltransferase (GGT) and creatine kinase (CK) assays were 13 7%, -13 9 and -19 2% respectively The precision of the enzyme assays were improved by using calibrators Conclusions The traceability of ALT, AST, AMY and LD assays were reasonable Calibrating by using enzyme calibrator can improve the assay precision among laboratories