Quality control system for automatic sediment analyzer
- VernacularTitle:对全自动尿液沉渣分析仪检测结果的质量控制
- Author:
Yingmu CAI
;
Jinghua LIN
;
Xiaoyang JIAO
- Publication Type:Journal Article
- Keywords:
Urine sediment;
Quality control
- From:
Chinese Journal of Laboratory Medicine
2001;0(03):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To establish a quality control process for Sysmex UF 100 Urine analyzer. Methods L J control process of UF 100 Urine analyzer; x m Q C methods; Chemistry and UF 100 methods comparison; Microscope examination Results L J control process can reflect and monitor the instrument status and the reagent quality; The x m Q C method can indicate the influence factor which come from the samples; Chemistry and UF 100′s comparison can find their contradiction and the error causes;Microscopy can make up UF 100′s test blind spots Conclusions These methods can monitor the result′s quality practically and effectively It is benefit to improve the test quality of UF 100 analyzer