Internal quality control method for routine biochemistry analysis
- VernacularTitle:生化常规质控方法探讨
- Author:
Ping LI
;
Henjian HUANG
;
Qingjiang ZHAO
- Publication Type:Journal Article
- Keywords:
Blood;
Biochemistry;
Internal guality control
- From:
Chinese Journal of Laboratory Medicine
2001;0(05):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To study the valid correct method for reagent selection, calibration of analyzers, and internal quality control protocol. Methods Reagents were selected by comparing with BRENGER Mannheim(BM)reagent used in Hitachi 7170A. Analyzers were calibrated once a week. One internal quality control chart was set up for three analyzers. Specimens were analyzed by the three machines respectively. Results Except for amylase, all the results obtained from the three analyzers were not significant. Conclusion The recommended quality control protocol is useful to obtain comparable results from different automatic biochemistry analyzers.