Study on P Wave Detection Method Based on Optimal Fitting Function
- VernacularTitle:基于最优化函数拟合的P波检测方法的研究(英文)
- Author:
Xiaoli QUAN
;
Nanquan ZHOU
;
Shunren HU
- Publication Type:Journal Article
- Keywords:
P wave;
fitting function method;
optimal interval;
the least square method
- From:Space Medicine & Medical Engineering
2006;0(02):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To accurately detect ECG P wave,a P wave detection algorithm based on fitting function in the optimal interval is proposed and evaluated in this paper,for resolving very difficult problems.Such as differences of positions and figures in ECG P wave as well as easily suffered diversified interferences.Methods In this algorithm the quadratic function to fit the P wave with least square method in every interval was used,which was shifted in local range.Then the optimal fitting interval of P wave was found by comparing the error of fitting.Finally,the characteristic points of P wave were obtained with the fitting function to fit P wave in the optimal interval.Results The performance of this algorithm tested with the records of the MIT-BIH database was effective and accurate.Conclusion This algorithm on the wide range of heart rate variation and small P wave of ECG P wave detection has good effect.Also it has some capabilities of anti-interference and quite low dismissal probability in particular.