Reliability study of 6-axis model surgery simulator for orthognathic surgery.
10.5125/jkaoms.2010.36.1.23
- Author:
Jae Ho JEON
1
;
Hyung Chul LEE
;
Hyun Jin JI
;
Yeong Jin JEON
;
Yong Il KIM
;
Woo Sung SON
;
Soo Byung PARK
;
Sung Sik KIM
;
Dae Seok WHANG
Author Information
1. Department of Orthodontics, School of Dentistry, Pusan National University, Korea. wsson@pusan.ac.kr
- Publication Type:Original Article
- Keywords:
Orthognathic surgery procedure;
Orthognathic surgery
- MeSH:
Orthognathic Surgery;
Plastics;
Splints
- From:Journal of the Korean Association of Oral and Maxillofacial Surgeons
2010;36(1):23-27
- CountryRepublic of Korea
- Language:Korean
-
Abstract:
The purpose of this study was to evaluate the reliability of 6-axis model surgery simulator (6AMSS) for orthognathic surgery. A rectangular parallelepiped plastic block was assembled to model-mounting plate of 6AMSS. Left-right (X), anterior-posterior (Y), up-down (Z) translation and pitching (empty set X), rolling (empty set Y) and yawing (empty set Z) rotation was planned and performed using 6AMSS. The actual translation and rotation were measured with dial gauge and precisional protractor, respectively. Comparison between the planned and actual movements of plastic block for each variable were made using paired t- test. Statistical analysis for X, Y, Z, empty set X, empty set Y and empty set Z movement have shown no significant differences between planned and actual movement (P > 0.05). This indicate that model surgery performed with the aid of the 6AMSS is accurate in 3D translation and rotation. The 6AMSS is practically useful for accurate fabrication of surgical splint for orthognathic surgery.