The Comparison of the KAPM Dosimetric Protocol (1990) with the TG-21 and Clambda/C(E) Method.
- Author:
Byong Yong YI
1
;
Eun Kyung CHOI
;
Hyesook CHANG
Author Information
1. Department of Therapeutic Radiology, Asan Medical Center, College of Medicine, University of Ulsan, Seoul, Korea.
- Publication Type:Original Article
- Keywords:
Dosimetric Protocol;
KAPM Dosimetric Protocol;
TG-21
- MeSH:
Water
- From:Journal of the Korean Society for Therapeutic Radiology
1991;9(2):337-342
- CountryRepublic of Korea
- Language:English
-
Abstract:
The Comparison of the KAPM Dosimetric Protocol (1990) with the TG-21 and Clambda/C(E) (ICRU-21 and SCRAD protocol) method is studied. The therapetutic range of radiation (photon 4MV-15MV and electron 6MeV-20MeV)and three kinds of the chambers were used in the water phantom. The Results from TG-21 and KAPM Protocol did not show much differences (less than 1%) throughout the whole energy range; N(D) from KAPM protocol and Ngas from TG-21 showed 0.2% deviation mainly from W/e difference between two protocols. But the results from KAPM protocol (1990) and those from Clambda/C(E) Method showed -1.9+/-0.6%(KAPM protocol is higher) deviation for photom beam and +3.3+/-1% (KAPM protocol is lower) deviation for electron beams.