Evaluation on application of electrolyte module in OLYMPUS AU 5421 Automatic Biochemical Analyzer
10.3969/j.issn.1673-4130.2014.09.038
- VernacularTitle:OLYMPUS AU5421型全自动生化分析仪电解质模块应用的评价
- Author:
Jie PAN
;
Xunyang LUO
;
Xiaohan HU
;
Kui ZHANG
- Publication Type:Journal Article
- Keywords:
precision;
accuracy;
linear range;
biochemical analyzer;
cross-contamination
- From:
International Journal of Laboratory Medicine
2014;(9):1177-1178,1180
- CountryChina
- Language:Chinese
-
Abstract:
Objective To evaluate the performance of electrolyte module in OLYMPUS AU5421 Automatic Biochemical Analy-zer in K+ ,Na+ ,Cl- detection .Methods Electrolyte module in OLYMPUS AU5421 Automatic Biochemical Analyzer was employed to detect K+ ,Na+ ,Cl- ,and their precision ,accuracy ,linearity range ,cross-contamination and the test results of instruments were compared .The linear regression equation and regression coefficients were calculated according to the methods of Clinical and Labo-ratory Standards Institute(CLSI) EP6-A ,EP9-A2 documents .Results The within-run precisions of K + ,Na+ ,Cl- were 0 .328% , 0 .306% ,0 .343% ,respectively ,and their between-run precisions were 0 .780% ,1 .012% ,0 .985% ,respectively .Their regression coefficients of linear equation were all greater than 0 .999 ,and their cross-contamination rate were 1 .066% ,0 .812% and 0 .858% , respectively .Conclusion Electrolyte module in OLYMPUS AU5421 Automatic Biochemical Analyzer modules has advantages of good precision ,accuracy ,linear range ,and low cross-contamination rate .