Animal experimental research of the endothelialization of home-made atrial septal defect occluder device
- VernacularTitle:国产房间隔缺损堵塞装置内皮化的实验研究
- Author:
Mingwu CHEN
;
Aiqing ZHOU
;
Fen LIN
- Publication Type:Journal Article
- Keywords:
Atrial septal defect;
Nitinol;
Endothelialization;
Medical device design
- From:
Journal of Interventional Radiology
2001;0(05):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To evaluate the endothelialization of Chinese nitinol atrial septal defect occlud-er device. Methods Atrial septal defect with controllable size was created by the Brockenborough needle and Rashkind balloon atrial septostomy, the occluder devices were implanted in six piglets (mean weight 7. 5 kg). Two pigs were killed each time after 1 month, 3 months and 6 months after the device implantation and then the explanted devices were examined by scanning electron microscope (SEM). Results The devices were found covering with collagen fibrosis together with diffuse endothelial cells spreading over the primer 1 month after implantation. The implants were covered mostly by ncointima 3 months after implantation and completely covered by confluent endothelial cells 6 months after the implantation. Endothelial cells were not found on the smooth marker band at 3 months, however, did exist by 6 months. Conclusions Home-made atriatseptal defect occluder devices were mostly endothelialised 3 months after the implantation and did completely at 6months.