Enzyme electrode for the determination of sodium lactate in sodium lactate injection
- VernacularTitle:酶电极法测定乳酸钠注射液中L-乳酸钠的含量
- Author:
Shiqing SUN
;
Ning WANG
;
Shulan GE
- Publication Type:Journal Article
- Keywords:
sodium L-lactate;
enzyme electrode;
biosensor;
sodium lactate injection
- From:
Chinese Pharmaceutical Journal
1998;(4):239-
- CountryChina
- Language:Chinese
-
Abstract:
OBJECTIVE: To determine the concentration of sodium L-lactate in sodium lactate injection. METHODS: Lactic oxidase was immobilized at a hydrogen peroxide electrode and the enzyme electrode was used for the amperometric determination of sodium L-lactate in sodium lactate injection. Sodium D, L-lactate in injection was analyzed by that technique according to Chinese Pharmacopoeia. RESULTS: The quantity of sodium L-lactate was about 10% in total amount of sodium D, L-lactate in sodium lactate injection. CONCLUSION: Sodium L-lactate in sodium lactate injection could be determined specially, rapidly and inexpensively.