Comparison on two methods of HPLC-MS/MS and MEIA in monitoring the tacrolimus concentrations for organ transplantation patients
- VernacularTitle:HPLC-MS/MS法与MEIA法在监测器官移植患者他克莫司全血浓度中的比较
- Author:
Pengfei LI
;
Lihong LIU
;
Ping MA
;
Chunlei DING
;
Weihang TONG
- Publication Type:Journal Article
- Keywords:
HPLC-MS/MS;
microparticle enzyme immunoassay;
organ transplantation;
tacrolimus
- From:
Chinese Journal of Clinical Pharmacology and Therapeutics
2000;0(01):-
- CountryChina
- Language:Chinese
-
Abstract:
AIM: To compare two methods of HPLC-MS/MS and MEIA to monitor the tacrolimus concentrations for organ transplantation patient.METHODS: After developing and validating the HPLC-MS/MS method to determine the concentrations of tacrolimus,the tacrolimus samples were quantitated by two methods of HPLC-MS/MS and MEIA,respectively.Evaluate two methods by statistical analysis of quantitative results.RESULTS: The mean concentration were(4.86?0.46) ng/mL by HPLC-MS/MS and(5.52?0.43) ng/mL by MEIA.The coefficient correlation of two methods was 0.8771,and the two methods have good correlations in statistics.The mean concentration ratio was(90.3?5.3)%.CONCLUSION: The HPLC-MS/MS method is a more precise method for to determining the effective concentration of tacrolimus,which is suitable for daily TDM.