Experimental Study in Pigs Using the Home-Made Stent(Shape Memory Alloy)
- VernacularTitle:自制金属内支架的动物实验研究
- Author:
Xiaolin WANG
;
Jiemin CHENG
;
Zhiping YAN
- Publication Type:Journal Article
- Keywords:
Memory alloy;
Stent;
Animal Experiment
- From:
Journal of Interventional Radiology
1994;0(03):-
- CountryChina
- Language:Chinese
-
Abstract:
Purpose:Home-made NiTi-stent was used in the study to observe the biological compatibility.Materials card Methods:Stents were implanted within the bile ducts,arteries,veins,bronchi of swine by surgery.Swine was killed according to the tie schedule to observe the patency of stents.Results:6 of 11 biliary stents were occluded completely,with only 5 stents partially patent.The degree of occlusion was related to the time period.10 of 11 femoral stents were patent,only 1 stent was occluded completely.1 of the only 1 venous stent was occluded completely,4 of 8 bronchial stants were patent.Epithlium tissues were found along the surfaces of stents.Both sides of the stents were covered by the epithelium cells partially and dominated by the proliferation of connective tissues and parenchymal cells of the organs.The constitution of proliferated tissues was related to the implanted time period of the stents.No connective tissues were found within two weeks,but obvious proliferation of connective tissues were found associated with lympheytic tissues.Conclusion:The dagree of surface covering of the home- made stent by epithelial the rate of was related to the diameters of the of and the lumceh stent implantation segment also the flow volume within the stent,and the period of stent implantation.Good results can be yielded by selecting the suitable stent and the implanting site.