Identification of hematuria by combination of Sysmex UF-50 cytometer of blood analyzer
- VernacularTitle:尿沉渣分析仪与血液分析仪联合检测鉴别血尿来源
- Author:
Jiyin QIAN
- Publication Type:Journal Article
- Keywords:
Hematuria;
Urinalysis;
Mean corpuscular volume
- From:
Chinese Journal of Primary Medicine and Pharmacy
2005;0(11):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To evaluate the value of combined detection for distinguishing glomerular hematuria from nonglomerular hematuria by Sysmex UF-50 automated flow cytometer with blood analyzer.Methods Mean corpuscular volume(MCV) of urinary red cell(RBC) in 414 urine samples from hematuria patients were measured by blood analyzer and all kinds of laser parameters of urinary RBC were measured by Sysmex UF-50 antomated flow cy- tometer,which were compared with the morphological results of RBC observed by light microscope.Results The RBC-Pv0FSC,RBC-MFse and RBC-Fsc-DW of hematuria with glomerular diseases and non-glomerular diseases were respectively (59.5?12.5)ch and (114.5?9.4)ch (P