Concentration of quinolinic acid in rat serum by GC-MS
- VernacularTitle:大鼠血清中喹啉酸含量的气质联用测定
- Author:
Ying XIONG
;
Huarong YU
;
Qixin ZHOU
- Publication Type:Journal Article
- Keywords:
GC-MS;
quinolinic acid;
serum
- From:Journal of Third Military Medical University
1983;0(04):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To establish a sensitive and accurate gas chromatography-mass spectrometry method for determination of quinolinic acid in rat serum.Methods The objective compound was purified and derivatized to its dihexafluoroisopropyl ester before quantitative determination by scanning method and quantitative determination by SIM method.Results The endogenous substances in serum had no interference on the measurement of quinolinic acid in the sample.Assay linearity was obtained in the range of 0.05-20 ?mol/L with limit of quantitation(LOQ)of 0.05 ?mol/L.The intra-day relative standard deviation(RSD)was lower than 15%.The inter-day relative standard deviation was lower than 12%.Conclusion It is a sensitive,accurate and convenient method.The dihexafluoroisopropyl ester of quinolinic acid is quite stable,but the serum should not be preserved for a long time before derivation.