Determination of somatostatin in serum by large pore liquid chromatography column
- VernacularTitle:大孔径液相色谱柱检测血清奥曲肽
- Author:
Fangnan LIU
;
Xiuzhen WAN
- Publication Type:Journal Article
- Keywords:
High performance liquid chromatography;
Somatostatin
- From:
Parenteral & Enteral Nutrition
2004;0(05):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective: To develop a method in the determination of somatostatin level in blood for clinical medication. Methods: A method for the determination of somatostatin(Octreotide) in serum by high performance liquid chromatography with diode-array detector(DAD) was developed.After ultrafiltration,the pH values of blood samples were adjusted to 4,and samples were analyzed by Kromasil C18 column(4.6 mm? 250 mm,5 ?m i.d.) with a large pore(300) by acetonitrile/water/trifluoroacetic acid(20/80/0.01%,v/v) as mobile phase. Results:The recovery rate of Octreotide was 85%,and the calibration curve showed good linearity in the range of 0.006-0.5 mg/mL. Conclusion: The method is suitable for the detection of Octreotide content by liquid chromatography.