Accuracy verification of PET-CT image fusion and its utilization in target delineation of radiotherapy
- VernacularTitle:PET-CT同机融合精度验证及在放疗靶区勾画中的应用
- Author:
Xuetao WANG
;
Jinming YU
;
Guoren YANG
;
Heyi GONG
- Publication Type:Journal Article
- Keywords:
Tomography,positron-emission;
Image processing,compute d-assisted;
Target delineations
- From:
Chinese Journal of Radiation Oncology
1993;0(03):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective Evaluate the accuracy o f co-registration of PET and C T (PET-CT) images on line with phantom, and utilize it on patients to provide c l inical evidence for target delineation in radiotherapy. Methods A phantom with m arkers and different volume cylinders was infused with various concentrations of 18FDG, and scanned at 4?mm by PET and CT respectively. Aft er having been transmi tted into GE eNTEGRA and treatment planning system (TPS) workstations, the image s were fused and reconstructed. The distance between the markers and the errors were monitored in PET and CT images respectively. The volume of cylinder in PET and CT images were measured and compared by certain pixel value proportion deduc tion method. The same procedure was performed on the pulmonary tumor image in te n patients. Results eNTEGRA and TPS workstations had a good length linearity, b ut the fusion error of the latter was markedly greater than the former. Tumors i n different volume filled by varying concentrations of 18FDG required different pixel deduction proportion. The cylinder volume of PET and CT images were almost the same, so were the images of pulmonary tumor of ten patients. Conclusions T he accuracy of image co-registration of PET-CT on line may fulfill the clinica l demand. Pixel value proportion deduction method can be used for target delineati on on PET image.