Assessment of lung volume using multi-slice spiral CT and comparison with pulmonary function test
- VernacularTitle:多层螺旋CT肺容积评估及其与肺功能的相关性研究
- Author:
Ken CHEN
;
Liguang ZOU
- Publication Type:Journal Article
- Keywords:
multi-slice spiral CT;
lung volume;
pulmonary function test;
three-dimension
- From:Journal of Third Military Medical University
2003;0(09):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To explore the value of multi-slice spiral CT (MSCT) with three dimensional (3D) postprocessing for assessing the lung volume and compare it with pulmonary function test (PFT). Methods In 60 cases, there were 34 healthy controls and 26 patients with emphysema. CT scan and PFT were performed within 3 d in all case. The total lung was respectively scanned at the end of full inspiration and full expiration, using GE Lightspeed 16 slices spiral CT. The total lung volumes and voxel index (VI) at threshold range of -1023 Hu to -900 Hu were measured by pulmonary quantitative software of AW4.2 workstation. Results The full expiration volume(Vex) significantly correlated with residual volume(RV) (r=0.908) as well as the full inspiration volume (Vin) with total lung capacity (TLC) (r=0.897). The value of (Vin-Vex) well correlated with forced vital capacity(FVC), as well as Vex/Vin with RV/TLC, VI Ain with TLC, VI Aex with RV. Conclusion Significant correlation of lung volume accessed by MSCT with 3D technique with PFT is useful for assessing the pulmonary function.