Effect of electromyogram variation on BIS accuracy during general anesthesia
- VernacularTitle:全身麻醉中肌电活动的变化对BIS监测准确性的影响
- Author:
Jing LIU
;
Weidong MI
;
Hong ZHANG
- Publication Type:Journal Article
- Keywords:
muscle relaxant;
anesthesia maintenance;
depth of anesthesia;
target-controlled infusion(TCI);
bispectral index(BIS);
electromyogram(EMG)
- From:
Medical Journal of Chinese People's Liberation Army
1982;0(01):-
- CountryChina
- Language:Chinese
-
Abstract:
40), the values of BIS, electromyogram (EMG), signal quality index (SQI), mean arterial pressure(MAP), and heart rate(HR) were recorded every minute for 5min, then additional vecuronium (0.05mg/kg) was administered, and the values of the parameters above mentioned were registered for the next 10min. Results BIS value decreased significantly 3 min after the administration of the additional vecuronium (P0.05). Conclusion A misleading BIS value may be caused by the decline of muscle relaxant during the maintenance stage of propofol/fentanyl anesthesia, which greatly influence the accuracy of BIS monitoring.