Application study on LC-ESI--MS/MS determination of inosine in human plasma
10.3969/j.issn.1671-8348.2014.03.020
- VernacularTitle:LC-ESI-M S/M S方法测定人血浆肌苷的应用研究
- Author:
Hongbo WANG
;
Mengliang ZHANG
;
Yunhui ZHANG
;
Yan YANG
;
Baochang ZHANG
- Publication Type:Journal Article
- Keywords:
inosine;
inosine pranobex;
liquid chromatography-tandem mass spectrometry;
plasma;
pharmacokinetics
- From:
Chongqing Medicine
2014;(3):311-313,316
- CountryChina
- Language:Chinese
-
Abstract:
Objective To develop a LC-MS/MS method for the determination of inosine in human plasma and to apply it in the isoprinosine pharmacokinetic researches .Methods With adefovir as the internal standard ,methanol -10 mmol/L ammonium ace-tate(15 :85 ,v/v) was adopted as the mobile phase .The chromatographic column was the Agilent SB-C18 column(5 μm ,150 mm × 4 .6 mm I .D .) .Electrospray ionization(ESI) source was applied and the detection was conducted by the multiple-reaction monito-ring(MRM) mode .The ion reactions for the quantitative analysis were m/z 135 .0-267 .3(inosine) and m/z 134 .1-272 .0(adefo-vir) .Results The linear range of plasma inosine was 10-3 000 ng/mL .The lower detection limit was 10 ng/mL .The inter-and in-tra-day precisions(RSD) were less than 5% .The average recovery rate was above 90% without the matrix effect .Conclusion This method has the strong specificity ,convenience in treating the sample and high sensitivity ,and is suitable for the inosine pharmacokinetic re-searches .