A NEW METHOD OF NON-INVASIVE DETERMINATION OF CARDIAC OUTPUT BY DynaPulse: COMPARISON WITH DIRECT FICK METHOD AND DOPPLER ECHOCARDIOGRAPHY
- VernacularTitle:DynaPulse动态血压仪测量心输出量与Fick法和多普勒超声法的比较
- Author:
Guorong XI
;
Tiande LI
;
Guan ZHI
- Publication Type:Journal Article
- Keywords:
DynaPulse;
echocardiagraphy, Doppler;
Fick method;
cardiac output;
measurement techniques
- From:
Medical Journal of Chinese People's Liberation Army
1981;0(04):-
- CountryChina
- Language:Chinese
-
Abstract:
DynaPulse, a newly developed noninvasive apparatus, is valuable in assessing blood pressure, cardiac output and vascular compliance. The purpose of the present study was to evaluate the validity of cardiac output measurement using DynaPulse in patients, compared with echocardiography and Fick method. Forty five inpatients underwent cardiac output measurement using DynaPulse (CO DP ) and Doppler echocardiography (CO DE ), cardiac output was measured with DynaPulse (CO DP ) and Fick method (CO FM ) in additional 26 patients. A good correlation was found between CO DP and CO DE ( r =0 76). There was an acceptable correlation between CO DP and CO FM ( r =0 61). DynaPulse can provide a noninvasive, clinically useful estimation of cardiac output.