Single-dose contrast-enhanced magnetic resonance aortography
- VernacularTitle:单倍剂量增强MR主动脉造影
- Author:
Bo JIANG
;
Quanfei MENG
;
Shenping YU
- Publication Type:Journal Article
- Keywords:
Magnetic resonance angiography;
Aortography;
Comparative study
- From:
Chinese Journal of Radiology
2001;0(07):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To probe the feasibility of single dose contrast enhanced magnetic resonance aortography(CEMRA) and evaluate its value in clinical application. Methods Transit time(TT) test and duration of peak enhancement of aorta(DPE) test in 28 healthy adults were performed. With 2 different injection rate of Gd DTPA at 2 ml/s and 3 ml/s, the TT,peak signal intensity of aorta(SPE),DPE and the interval between DPE and the time of injection(?T) were measured and compared. Two groups of healthy volunteers(each 10) underwent three diamentional fast imaging in static procession sequence (3D FISP) CEMRA with standard dosage of Gd DTPA with the injection rate at 2 ml/s and 3 ml/s respectively. The signal/noise ratio(SNR) of the source images and the image quality of CEMRA of both groups were compared. Fifteen patients with aortic disease did the same study as the voluteers with the injection rate only at 3 ml/s. Results With the injection rate of 2 ml/s and 3 ml/s,the TT,SPE,DPE,?T and the SNR of CEMRA in two groups were significantly different( P