An experimental study on the embolus-capturing efficiency of a new-designed retrievable artery filter
- VernacularTitle:一种可取出式动脉滤器的实验研究
- Author:
Long JIN
;
Yinghua ZOU
- Publication Type:Journal Article
- Keywords:
Thromboembolism;
Protective devices;
Radiology, interventional;
Models, cardiovascular
- From:
Chinese Journal of Radiology
1994;0(06):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To develop a physical in vitro model of the human carotid artery(CA) and analyze the factors that influence the embolus-capturing efficiency of the retrievable artery filters. Methods The flow model closely simulated the physiological parameters of the human carotid artery. The influence of the following experimental parameters on the capturing rate was analyzed with a multiple logistic regression model: embolus diameter and length, CA diameter, flow quantity, type of filter, and type of testing. Results At the 5% level,except for flow quantity,a significant influence on the capturing rate could be demonstrated for the 6 factors mentioned above, the embolus diameter was the strongest influencing factor. In CA with diameters of 4,5,6,7 mm, the 8-wire filter captured (90?7)%, (91?8)%, (81?8)%,and (68?9)% of 0.5 mm emboli and (93?4)%, (95?6)%, (95?6)%,and (82?11)% of 1 mm emboli. In all, the smaller, the interlacing opening of the filter, the higher the capturing rate, and the presence of an embolus in a filter would decrease the capturing rate for the second embolus. Conclusion Retrievable artery filter is a promising cerebral protection device. The physical in vitro model is an effective alternative in studying the influence factors of the capturing rate of the filters.