Influence of mandibular metallic prothesis to the radiation dose in the neighboring tissue: An experiment study.
- VernacularTitle:金属下颌骨修复板对界面组织放射剂量分布影响的研究
- Author:
Xiaoli CHEN
;
Zhiyuan ZHANG
;
Zhonghe WANG
- Publication Type:Journal Article
- Keywords:
Co ?-ray Radiation dose distribution Mandibular metallic prothesis
- From:
Chinese Journal of Radiation Oncology
1992;0(01):-
- CountryChina
- Language:Chinese
-
Abstract:
Purpose: To measure the distribution of radiation dose on neighboring tissue of the mandible metallic. Materials and Methods: The stainless steel and tantalum prothesis of mandible were inserted in the phantoms. The percent radiation dose distributions of 60 Co ?-ray were measured by Lithium fluoride thermo-luminsence method. Results: Using unidirectional beam, the radiation dose was increased in the front of metallic and decreased at the back of it in 3mm thickness. Stainless steel gave higher changes than tantalum. Using opposite beams irradiation, these changes were compensated each other. Conclusion: Tantalum prothesis is advised to mandible when 60 Co is used.