Determination of Aluminium in Drinking Water by Flow Injection Analysis Instrument
- VernacularTitle:饮用水中铝的流动注射分析测定法
- Author:
Yuting LIU
;
Li ZHU
- Publication Type:Journal Article
- Keywords:
Water;
Flow injection analysis;
Aluminium
- From:
Journal of Environment and Health
2007;0(11):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To establish a method for the determination of aluminium in drinking water by flow injection analysis instrument.Methods In a hexamine buffer solution(pH=6.0),phenanthroline solution and ascorbic acid as the anti-interference agent,aluminium could react with chromazurol S solution and form a compound,of which maximum absorption was at 550 nm.Results The linear range for this method was 5-200 ?g/L.The correlation coefficient was 0.999 3,the detection limit of this method was 2.0 ?g/L,the recovery rates were in the range of 92.2%-108% and RSD were in the range of 2.5%-8.2%.Conclusion The flow injection method is simple,rapid,accurate and is applicable to the determination of aluminium in drinking water.