Intradermal stimulus and sensitivity response for new kinds of Mg alloy
- VernacularTitle:新型镁合金皮内刺激与致敏实验研究
- Author:
Fei WANG
;
Gang WANG
;
Xiangxiang MENG
- Publication Type:Journal Article
- Keywords:
Mg alloy;
stimulus;
sensitivity;
biomaterial;
step anodization
- From:
Orthopedic Journal of China
2006;0(13):-
- CountryChina
- Language:Chinese
-
Abstract:
[Objective] To evaluate the biocompatibility in vivo of Mg alloy AZ91D(AZ0),Mg-Mn-Ce(RE0)and their step anodizational materials(AZ1,AZ2,AZ3,RE1,RE2,RE3)by the experiments of intradermal stimulus and sensitivity response.[Methods]According toGB/T 16886.10-1997 Biological Evaluation of Medical Apparatus and Instrument,compared with normal sodium,pure magnesium and medical Ti-6Al-7Nb,the leaching liquor were injected into cutis to evaluate stimulus response,to evaluate sensitivity response by injection and illinition.[Results]RE2 could not prepare leaching liquor,it's kicked out.Intradermal stimulus response:normal sodium,Ti-6A1-7Nb,AZ1,AZ3 groups inflammation responded as degree I,scored 0.The others inflammation responded as degree Ⅱ,scored 1.Sensitivity response:formaldehyde group inflammation responded as degree IV,erythema and edema were degree IV.Normal sodium group inflammation responded as degree I,erythema and edema were degree 0.Ti-6Al-7Nb,AZ1,AZ3 groups inflammation responded as degree Ⅱ,erythema and edema were degree 1.The others inflammation responses were degree Ⅲ,erythema and edema were degree Ⅱ.[Conclusion]The new kinds of magnesium alloy AZ1,AZ3 possesses good preliminary biocompatibility in vivo,so it is very possible to be a new type of bone surgery implant material.