Study and biomechanical evaluation of the device of gear-distraction plate for spine reduction and fixation
- VernacularTitle:齿轮撑开式脊柱复位固定板装置的研制及生物力学评价
- Author:
Xuewen CUI
;
Lianshun JIA
;
Wen YUAN
- Publication Type:Journal Article
- Keywords:
gear distraction plate (GDP);
fixation plate;
internal fixation;
biomechanics
- From:
Orthopedic Journal of China
2006;0(20):-
- CountryChina
- Language:Chinese
-
Abstract:
[Objective]A new device of gear-distraction plate (GDP) for spine reduction and fixation was designed and its biomechanical characteristics was evaluated.[Method]The GDP implants were made of titanium alloy (TC4,Ti6AL4V) and the instruments were made of stainless steel after design.Eighteen fresh calf lumbar specimens were randomly divided into three groups, the GDP group had a biomechanical test contrast to control group(CD, Steffee) in load-responsive change, load-displacement, strength, stiffness, torsion intensity and ultimate strength.[Result]Results It is better than control group in load-responsive change, load-displacement, strength, stiffness, torsion intensity and ultimate strength, there was a significant difference between them (P