Design and clinical application of anatomical locked hook-plate and internal fixation system for femoral intertrochanteric fracture
- VernacularTitle:解剖型股骨粗隆锁定钩板内固定系统的设计制作与临床应用
- Author:
Yunfeng LI
;
Baiwei LIU
;
Jian LU
- Publication Type:Journal Article
- Keywords:
femur;
femoral intertrochanteric fracture;
anatomical;
locked hook-plate;
internal fixation
- From:
Orthopedic Journal of China
2006;0(14):-
- CountryChina
- Language:Chinese
-
Abstract:
[Objective]To explore a new type of internal fixation system for the fixation of femoral intertrochanteric fracture.[Method]According to femoral tuberosity anatomical shape,anatomical locked hook-plate internal fixation system(ALHP) was designed,its biomechanical properties was examined and ALHP was applied for the clinical treatment of intertrochanteric fracture of the aged patients.[Result]Twenty-seven cases of intertrochanteric fracture were fixed with ALHP and followed up from 1.5 to 2 years,with an average time was 1.8 years.According to Sanders functional sores:the good and excellent rate was 100%.[Conclusion]Anatomy type locked hook-plate internal fixation system possesses reasonable design and produces firm fixation for intertrochanteric fracture.It can offer the elderly patients early exercises and reduce the incidence of complication.