Determination of Tellurium in Workplace Air by Microwave Digestion-HG-AFS
	    		
		   		
	    	
    	
    	
   		
        
        	
        		- VernacularTitle:空气中碲的微波消解-氢化物发生原子荧光测定法
 
        	
        	
        	
        		- Author:
	        		
		        		
		        		
			        		Lijun YANG
			        		
			        		;
		        		
		        		
		        		
			        		Shusong JIA
			        		
			        		;
		        		
		        		
		        		
			        		Weijing GUO
			        		
			        		
		        		
		        		
		        		
		        		
		        		
			        		
			        		
		        		
	        		
        		 
        	
        	
        	
        		- Publication Type:Journal Article
 
        	
        	
        		- Keywords:
        			
	        			
	        				
	        				
			        		
				        		Spectrophotometry;
			        		
			        		
			        		
				        		Microwave digestion;
			        		
			        		
			        		
				        		Hydride generation atomic fluorescence spectrometry;
			        		
			        		
			        		
				        		Tellurium
			        		
			        		
	        			
        			
        		
 
        	
            
            
            	- From:
	            		
	            			Journal of Environment and Health
	            		
	            		 1992;0(02):-
	            	
            	
 
            
            
            	- CountryChina
 
            
            
            	- Language:Chinese
 
            
            
            	- 
		        	Abstract:
			       	
			       		
				        
				        	Objective To establish a method for determination of tellurium(Ti)in the air of workplace by microwave digestion-HG-AFS. Methods The air samples were collected by micropore filters and the concentration of tellurium was determined by microwave digestion-HG-AFS. Results In this method, the linear range was 0.11-40.0 ?g/L, the detection limit was 0.11 ?g/L, the lowest detected concentration of tellurium was 0.000 09 mg/m3(based on 30 L air sample), the rate of recovery was in the range of 95.5%-98.2%, the RSD was less than 3%. Conclusion The method is rapid, accurate, sensitive and suitable for the determination of tellurium in the air of workplace.