The Relationship between the Peak Value of Contrast Media Reaching Abdominal Aorta and Patients’ Factors
- VernacularTitle:造影剂到达腹主动脉的峰值大小与患者因素的关系
- Author:
Xiaomei ZHU
;
Dehang WANG
;
Mei LI
;
Lijun TANG
- Publication Type:Journal Article
- Keywords:
abdominal aorta;
peak value;
tomography ,X-ray computed
- From:
Journal of Practical Radiology
1991;0(03):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To investigate the relationship between the peak value of contrast media reaching abdominal aorta andpatients’ factors.Methods Ominpaque 300 mgI/ml 20 ml was injected at a rate of 2.5 ml/s in 108 patients and after 12 s was tracked by test bolus in the abdominal aorta at the level of celiac trunk, then the peak value in the abdominal aorta was achieved by the software dynamic evaluation. Regression analysis and stepwise regression analysis were used to study how the peak value of contrast media reaching abdominal aorta influenced by the patients’ factors including patients’ gender, age, height, weight, heart rate, blood pressure, injection site, peak time, and whether being with heart disease , diabetes mellitus or chemo . Results The peak value of contrast media reaching abdominal aorta was lower in male than that in female averagely, decreased gradually as the patients’ height, weight and peak time increased. The peak value of the injection site in the vein of hand back was lower than that of the cubitus generally and it was not influenced by the patients’ heart rate, blood pressure or whether being with heart disease, diabetes mellitus or chemo. Recommended formula of peak value(HU)=383.8400-Height(cm)?1.0909-Weight(kg)?0.6760+injection site?16.7878-peak time(s)?1.6882.Conclusion The volume of contrast media used in patients needing exam of CT angiography can be adjusted according to the patients’ age, gender, height, weight, injection site and peak time.