Determination of Dioxin in Wastewater by Pretreatment of Solid Phase Microextraction
- VernacularTitle:废水中二叶恶英类化合物测定的固相微萃取前处理法
- Author:
Yanqun LIU
;
Yikai ZHOU
;
Shu REN
- Publication Type:Journal Article
- Keywords:
Laboratory techniques and procedure: Solid phase microextraction: Wastewater: Dioxin
- From:
Journal of Environment and Health
1993;0(01):-
- CountryChina
- Language:Chinese
-
Abstract:
Objective To explore the determination of dioxin in wastewater samples pretreated by solid phase microextraction(SPME). Methods The wasterwater samples were purified and enriched by SPME, then were determined with HRGC-HRMS for the concentration of dioxin in wastewater samples. Results The most suitable conditions of the pretreatment were 30 min microextraction at 45 ℃. The detection limit was 0.05 pg/?l. RSD was lower than 10%. The recovery rates were 99%-102%. The concentration of total dioxins in wastewater sample was 0.78 pg/?l. Conclusion The method of the determination of dioxins in wastewater by SPME was simple and quick, and presented a broad prospects for application.