The Development of Quality Assurance Program for CyberKnife.
- Author:
Jisun JANG
1
;
Dong Han LEE
;
Young Nam KANG
;
Dong Oh SHIN
;
Moon Chan KIM
;
Sei Chul YOON
;
Ihl Bohng CHOI
;
Mi Sook KIM
;
Chul Koo CHO
;
Seong Yul YOO
;
Soo Il KWON
Author Information
1. CyberKnife Center, Kangnam St. Marys Hospital, Korea.
- Publication Type:Original Article
- Keywords:
CyberKnife;
Stereotactic radiotherapy;
Quality assurance program
- MeSH:
Constitution and Bylaws;
Humans;
Korea;
Mortuary Practice;
Quality Control;
Radiotherapy
- From:The Journal of the Korean Society for Therapeutic Radiology and Oncology
2006;24(3):185-191
- CountryRepublic of Korea
- Language:Korean
-
Abstract:
PURPOSE: Standardization quality assurance (QA) program of CyberKnife for suitable circumstances in Korea has not been established. In this research, we investigated the development of QA program for CyberKnife and evaluation of the feasibility under applications. MATERIALS AND METHODS: Considering the feature of constitution for systems and the therapeutic methodology of CyberKnife, the list of quality control (QC) was established and divided dependent on the each period of operations. And then all these developed QC lists were categorized into three groups such as basic QC, delivery specific QC, and patient specific QC based on the each purpose of QA. In order to verify the validity of the established QA program, this QC lists was applied to two CyberKnife centers. The acceptable tolerance was based on the undertaking inspection list from the CyberKnife manufacturer and the QC results during last three years of two CyberKnife centers in Korea. The acquired measurement results were evaluated for the analysis of the current QA status and the verification of the propriety for the developed QA program. RESULTS: The current QA status of two CyberKnife centers was evaluated from the accuracy of all measurements in relation with application of the established QA program. Each measurement result was verified having a good agreement within the acceptable tolerance limit of the developed QA program. CONCLUSION: It is considered that the developed QA program in this research could be established the standardization of QC methods for CyberKnife and confirmed the accuracy and stability for the image-guided stereotactic radiotherapy.