Determination of Nitrite in Water by High Performance Liquid Chromatography
- VernacularTitle:水样中亚硝酸盐的高效液相色谱测定法
- Author:
Jinzhi CHEN
;
Tie HAN
;
Qin ZHANG
- Publication Type:Journal Article
- Keywords:
High performance liquid chromatography;
Indirect UV;
Water sample;
Nitrite
- From:
Journal of Environment and Health
1993;0(03):-
- CountryChina
- Language:Chinese
-
Abstract:
ve To establish a method of high performance liquid chromatography (HPLC) for determi-nation of nitrite in water. Methods The trace content of nitrite in water sample was determined by indirect UV-HPLC. The mixture of methanol/o-phthalic acid (pH value was adjusted to 8.6 by 0.10 mol/L NaOH solution)5: 95 was defined as mobile phase. The water sample was directly filtered by chromatographic column pack with ODP. Results Under the conditions of wave length of 270 nm and flow rate of 0.9 ml/ min, the linear range of this assay was 0.0~20.0?g/ml nitrite, the relative standard deviation was 3.9%. The average recovery rate and detection limit were 99.9% and 0.001?g/ ml respectively. Conclusion This method could be applied to the determination of trace amount of nitrite in drinking water, purified water and mineral water.