Determination of Chromium (Ⅵ) in Waste Water by Inductively Coupled Plasma-Atomic Emission Spectrometry
- VernacularTitle:废水中六价铬的电感耦合等离子体发射光谱测定法
- Author:
Nanning SHI
;
Jianxin ZHANG
;
Luosheng XU
- Publication Type:Journal Article
- Keywords:
Inductively coupled plasma-atomic emission spectrometry;
Chromium (?);
Waste water
- From:
Journal of Environment and Health
1993;0(03):-
- CountryChina
- Language:Chinese
-
Abstract:
ve To study the new method for determination of chromium (Ⅵ) in waste water. Methods The chromium (Ⅵ) from waste water samples was seperated by ion-exchange method and concentrated, then was analyzed by inductively coupled plasma-atomic emission spectrometry. Results The precision, average recovery rate and detection limit were 2.05%, 99.7% and 0.018?g/ml respectively. Conclusion This method was easy and rapid to operate, and revealed less interference in determination of chromium (Ⅵ) in industrial waste water.