ANALYSIS ON INTRACARDIAC ELECTROGRAM SPECTRUM AND ITS APPLICATION IN CARDIAC PACING
- VernacularTitle:空内心电图频谱分析及其在心脏起搏技术中的应用
- Author:
Hong LIU
- Publication Type:Journal Article
- Keywords:
speetrum analysis;
pacemaker;
electrodes
- From:
Journal of Xi'an Jiaotong University(Medical Sciences)
1981;0(02):-
- CountryChina
- Language:Chinese
-
Abstract:
The spectrums of intracardiac electrograms, QRS complex wave and T wave which were picked up at the position where pacing electrode is to be located in the endocardia were investigated. Simultaneously, a variety of pacemaker sensitivity testing waves and the load capacities of two kinds of pacing electrode were analysised by a spectral analysis method. The results show that the sp(?)ctral distributions of intracardiac electrograms are below 70Hz, the T waves have lower spectral distributions and the QRS complexs have higher ones, indicating that the frequency band of the QRS wave detection circuit should be 8-45 Hz. The advantages and shortcomings of conventional pacemaker sensitivity testing waves and the load capacities of the pacing electrodes are discussed.